Optical Specifications & Displayed Information
Wavelength 600 nm
Photometric Range OD600: 0 – 4 A; McFarland: 0 – 16 MFU
Accuracy OD600: @ 1A: < ±0.01 A | McFarland: @ 0 – 8 MFU: ≤ ±0.1 MFU
Reproducibility OD600: @ 1A: < ±0.002 A | McFarland: @ 0 – 8 MFU: ≤ ±0.05 MFU
Cuvettes/Tubes DiluCells;
10 mm macro- and semi-micro cuvettes;
10, 12, 16, 18 mm glass/falcon tubes
– Without the need for adapters.
Wipeable sample compartment – fixed with drain hole.
General Specifications
Instrument Size 140 (L) x 110 (W) x 70 (H) mm
Weight Weight 0.6 kg
Display Size 3.5″ glove compatible touchscreen, 480 x 320 pixels
Memory 4 GB internal memory, access to all previously measured sample data in sample list feature
Time and Date Data Manual or automatic time setting
Connectivity USB A port for data export;
USB C for charging;
LIMS integration via WiFi, Ethernet/LAN (Rest API PUSH)
Battery Life Rechargeable, continuous use for ~8 hours
Certification CE, IEC61010-1:2012 and EN 61326-1:2013
Data Format .xls and .csv format exportable. For LIMS: JSON

Features and specifications are subject to change without notice

Compatible Sample Containers

Container Type Path Length Minimum Volume Virtual Dilution Adapter
DiluCell™ 10 0.9 mm 300 μl 1:11
Macro Cuvette 10 mm 1 ml
Semi-micro Cuvette 10 mm 0.5 ml
10 mm Tubes N/A* 0.9 ml
12 mm Tubes N/A* 1.1 ml
16 mm Tubes N/A* 2.2 ml
18 mm Tubes N/A* 2.2 ml

*Depends on tube being used. Features and specifications are subject to change without notice