Technical Specs

implen Diluphotometer cell density bacterial growth yeast growth

Implen Diluphotometer™

Implen OD600 cell density bacterial growth yeast growth

Implen OD600®

Main ApplicationsOptical density at 600 nm (OD600)Optical density at 600 nm (OD600) &
Turbidity standards (McFarland)
Photometric Accuracy< ±0.05 @ 1A

±0.01 @ 1A

 

±0.1 @ 0-8 McF

Photometric Range–0.3 – 1.99 A0 – 4 A
Photometric Repeatability< ±0.02 @ 1 A±0.002 @ 1A
±0.05 @ 0 – 8 MFU
Spectral Bandwidth40 nm18 nm
Optical Height8.5 mm8.5 mm
Operation

LCD Display

6 Buttons

3.5″ glove-compatible touchscreen

– easy and quick to clean

LanguagesEnglish

English, Chinese, French,

Spanish, Japanese, German

Compatible Vessel Types

10 mm macro- and semi-micro cuvettes

14, 16 mm round tubes

DiluCell™ 10

10 mm macro- and semi-micro cuvettes
10, 12, 16, 18 mm tubes

DiluCell™ 10
No adapter needed

Onboard Memory99 readings4 GB internal storage
> 500,000 measurements
Data ExportVia Software Tool;
Formats: Excel

USB 2.0 A – Flash Drive;

Formats: Excel and .csv (for LIMS: JSON)

ConnectivityRS232LIMS integration via WiFi or Ethernet/Lan
Internal Rechargeable BatteryYesYes
Dimensions180 x 150 x 60 mm140 x 110 x 70 mm
Weight0.6 kg0.6 kg
 No longer availablePurchase
Learn more about our new Implen OD600® device offering a higher concentration range, accuracy and reproducibility, featuring a 3.5″ glove-compatible display, user-friendly interface, utmost compatibility without the need for adapters, large data storage, universal capabilities, and effortless data export.